Experimental validation of Monte Carlo based treatment planning system in bone density equivalent media

  • Slaven Jurković UH RijekaMedical Physics Department
  • Đeni Smilović Radojčić
  • Božidar Casar
  • David Rajlić
  • Manda Švabić Kolacio
  • Ignasi Mendez
  • Nevena Obajdin
  • Dea Dundara Debeljuh

Abstract

Abstract:

 

Introduction: Advanced, Monte Carlo (MC) based dose calculation algorithms, determine absorbed dose as dose to medium-in-medium (Dm,m) or dose to water-in-medium (Dw,m). Some earlier studies identified the differences in the absorbed doses related to the calculation mode, especially in the bone density equivalent (BDE) media. Since the calculation algorithms built in the treatment planning systems (TPS) should be dosimetrically verified before their use, we analyzed dose differences between two calculation modes for the Elekta Monaco TPS. We compared them with experimentally determined values, aiming to define a supplement to the existing TPS verification methodology.

 

Results: Measured and calculated doses were coherent for water and lung equivalent density materials, regardless of the calculation mode. However, in the BDE part of the phantom, mean dose differences between the calculation options ranged from 5.73 to 8.31%, depending on the method used. In the BDE part of the phantom, neither of the two calculation options were consistent with experimentally determined doses.

Author Biography

Slaven Jurković, UH RijekaMedical Physics Department

Head of Medical Physics department

Published
2020-11-13
How to Cite
Jurković, S., Smilović Radojčić, Đeni, Casar, B., Rajlić, D., Švabić Kolacio, M., Mendez, I., Obajdin, N., & Dundara Debeljuh, D. (2020). Experimental validation of Monte Carlo based treatment planning system in bone density equivalent media. Radiology and Oncology, 54(4), 495-504. Retrieved from https://radioloncol.com/index.php/ro/article/view/3488
Section
Medical Physics